Arbitrary Waveform Generator
177
Press More 1/2 softkey to display the SWEEP menu page
2/2.
Softkey Options Description
Sweep
Mode
Up
Sweep from start frequency
to stop frequency.
Down
Sweep from stop frequency
to start frequency.
Up-Down
Sweep up and down between
start and stop frequencies.
It will sweep up first, and then
down.
Sweep
Time
----
Select and specify the sweep
time from start to stop
frequency.
Sweep
Ampl
----
Select and specify the
waveform amplitude.
Offset
----
Select and specify the offset
voltage.
More
2/2
---- Select page 1/2
Note: The sweep time specifies the number of seconds
required to sweep from the start frequency to the
stop frequency. The number of discrete frequency
points in the sweep is automatically calculated
according to the sweep time you select.